81_FR_62318 81 FR 62144 - Agency Information Collection Activities; Proposed Collection; Comment Request; Medical Device: Current Good Manufacturing Practice Quality System Regulations

81 FR 62144 - Agency Information Collection Activities; Proposed Collection; Comment Request; Medical Device: Current Good Manufacturing Practice Quality System Regulations

DEPARTMENT OF HEALTH AND HUMAN SERVICES
Food and Drug Administration

Federal Register Volume 81, Issue 174 (September 8, 2016)

Page Range62144-62148
FR Document2016-21553

The Food and Drug Administration (FDA) is announcing an opportunity for public comment on the proposed collection of certain information by the Agency. Under the Paperwork Reduction Act of 1995 (the PRA), Federal Agencies are required to publish notice in the Federal Register concerning each proposed collection of information including each proposed extension of an existing collection of information, and to allow 60 days for public comment in response to the notice. This notice solicits comments on recordkeeping requirements related to the medical devices current good manufacturing practice (CGMP) quality system (QS) regulation (CGMP/QS regulation).

Federal Register, Volume 81 Issue 174 (Thursday, September 8, 2016)
[Federal Register Volume 81, Number 174 (Thursday, September 8, 2016)]
[Notices]
[Pages 62144-62148]
From the Federal Register Online  [www.thefederalregister.org]
[FR Doc No: 2016-21553]


-----------------------------------------------------------------------

DEPARTMENT OF HEALTH AND HUMAN SERVICES

Food and Drug Administration

[Docket No. FDA-2016-N-2544]


Agency Information Collection Activities; Proposed Collection; 
Comment Request; Medical Device: Current Good Manufacturing Practice 
Quality System Regulations

AGENCY: Food and Drug Administration, HHS.

ACTION: Notice.

-----------------------------------------------------------------------

SUMMARY: The Food and Drug Administration (FDA) is announcing an 
opportunity for public comment on the proposed collection of certain 
information by the Agency. Under the Paperwork Reduction Act of 1995 
(the PRA), Federal Agencies are required to publish notice in the 
Federal Register concerning each proposed collection of information 
including each proposed extension of an existing collection of 
information, and to allow 60 days for public comment in response to the 
notice. This notice solicits comments on recordkeeping requirements 
related to the medical devices current good manufacturing practice 
(CGMP) quality system (QS) regulation (CGMP/QS regulation).

DATES: Submit either electronic or written comments on the collection 
of information by November 7, 2016.

ADDRESSES: You may submit comments as follows:

Electronic Submissions

    Submit electronic comments in the following way:

[[Page 62145]]

     Federal eRulemaking Portal: http://www.regulations.gov. 
Follow the instructions for submitting comments. Comments submitted 
electronically, including attachments, to http://www.regulations.gov 
will be posted to the docket unchanged. Because your comment will be 
made public, you are solely responsible for ensuring that your comment 
does not include any confidential information that you or a third party 
may not wish to be posted, such as medical information, your or anyone 
else's Social Security number, or confidential business information, 
such as a manufacturing process. Please note that if you include your 
name, contact information, or other information that identifies you in 
the body of your comments, that information will be posted on http://www.regulations.gov.
     If you want to submit a comment with confidential 
information that you do not wish to be made available to the public, 
submit the comment as a written/paper submission and in the manner 
detailed (see ``Written/Paper Submissions'' and ``Instructions'').

Written/Paper Submissions

    Submit written/paper submissions as follows:
     Mail/Hand delivery/Courier (for written/paper 
submissions): Division of Dockets Management (HFA-305), Food and Drug 
Administration, 5630 Fishers Lane, Rm. 1061, Rockville, MD 20852.
     For written/paper comments submitted to the Division of 
Dockets Management, FDA will post your comment, as well as any 
attachments, except for information submitted, marked and identified, 
as confidential, if submitted as detailed in ``Instructions.''
    Instructions: All submissions received must include the Docket No. 
FDA-2016-N-2544 for ``Agency Information Collection Activities; 
Proposed Collection; Comment Request; Medical Device: Current Good 
Manufacturing Practice Quality System Regulations.''
    Received comments will be placed in the docket and, except for 
those submitted as ``Confidential Submissions,'' publicly viewable at 
http://www.regulations.gov or at the Division of Dockets Management 
between 9 a.m. and 4 p.m., Monday through Friday.
     Confidential Submissions--To submit a comment with 
confidential information that you do not wish to be made publicly 
available, submit your comments only as a written/paper submission. You 
should submit two copies total. One copy will include the information 
you claim to be confidential with a heading or cover note that states 
``THIS DOCUMENT CONTAINS CONFIDENTIAL INFORMATION.'' The Agency will 
review this copy, including the claimed confidential information, in 
its consideration of comments. The second copy, which will have the 
claimed confidential information redacted/blacked out, will be 
available for public viewing and posted on http://www.regulations.gov. 
Submit both copies to the Division of Dockets Management. If you do not 
wish your name and contact information to be made publicly available, 
you can provide this information on the cover sheet and not in the body 
of your comments and you must identify this information as 
``confidential.'' Any information marked as ``confidential'' will not 
be disclosed except in accordance with 21 CFR 10.20 and other 
applicable disclosure law. For more information about FDA's posting of 
comments to public dockets, see 80 FR 56469, September 18, 2015, or 
access the information at: http://www.fda.gov/regulatoryinformation/dockets/default.htm.
    Docket: For access to the docket to read background documents or 
the electronic and written/paper comments received, go to http://www.regulations.gov and insert the docket number, found in brackets in 
the heading of this document, into the ``Search'' box and follow the 
prompts and/or go to the Division of Dockets Management, 5630 Fishers 
Lane, Rm. 1061, Rockville, MD 20852.

FOR FURTHER INFORMATION CONTACT: FDA PRA Staff, Office of Operations, 
Food and Drug Administration, 8455 Colesville Rd., COLE-14526, Silver 
Spring, MD 20993-0002, [email protected].

SUPPLEMENTARY INFORMATION: Under the PRA (44 U.S.C. 3501-3520), Federal 
Agencies must obtain approval from the Office of Management and Budget 
(OMB) for each collection of information they conduct or sponsor. 
``Collection of information'' is defined in 44 U.S.C. 3502(3) and 5 CFR 
1320.3(c) and includes Agency requests or requirements that members of 
the public submit reports, keep records, or provide information to a 
third party. Section 3506(c)(2)(A) of the PRA (44 U.S.C. 3506(c)(2)(A)) 
requires Federal Agencies to provide a 60-day notice in the Federal 
Register concerning each proposed collection of information, including 
each proposed extension of an existing collection of information, 
before submitting the collection to OMB for approval. To comply with 
this requirement, FDA is publishing notice of the proposed collection 
of information set forth in this document.
    With respect to the following collection of information, FDA 
invites comments on these topics: (1) Whether the proposed collection 
of information is necessary for the proper performance of FDA's 
functions, including whether the information will have practical 
utility; (2) the accuracy of FDA's estimate of the burden of the 
proposed collection of information, including the validity of the 
methodology and assumptions used; (3) ways to enhance the quality, 
utility, and clarity of the information to be collected; and (4) ways 
to minimize the burden of the collection of information on respondents, 
including through the use of automated collection techniques, when 
appropriate, and other forms of information technology.

Medical Devices: Current Good Manufacturing Practice Quality System 
Regulation--21 CFR Part 820--OMB Control Number 0910-0073--Extension

    Under section 520(f) of the Federal Food, Drug, and Cosmetic Act 
(the FD&C Act) (21 U.S.C. 360j(f)), the Secretary of the Department of 
Health and Human Services has the authority to prescribe regulations 
requiring that the methods used in, and the facilities and controls 
used for, the manufacture, preproduction design validation (including a 
process to assess the performance of a device, but not including an 
evaluation of the safety and effectiveness of a device), packing, 
storage, and installation of a device conform to CGMP, as described in 
such regulations, to assure that the device will be safe and effective 
and otherwise in compliance with the FD&C Act.
    The CGMP/QS regulation implementing authority provided by this 
statutory provision is found under part 820 (21 CFR part 820) and sets 
forth basic CGMP requirements governing the design, manufacture, 
packing, labeling, storage, installation, and servicing of all finished 
medical devices intended for human use. The authority for this 
regulation is covered under sections 501, 502, 510, 513, 514, 515, 518, 
519, 520, 522, 701, 704, 801, and 803 of the FD&C Act (21 U.S.C. 351, 
352, 360, 360c, 360d, 360e, 360h, 360i, 360j, 360l, 371, 374, 381, and 
383). The CGMP/QS regulation includes requirements for purchasing and 
service controls, clarifies recordkeeping requirements for device 
failure and complaint investigations, clarifies requirements for 
verifying/validating production processes and process or product 
changes, and clarifies requirements for product acceptance activities 
quality

[[Page 62146]]

data evaluations and corrections of nonconforming product/quality 
problems.
    Requirements are compatible with specifications in the 
international standards ``ISO 9001: Quality Systems Model for Quality 
Assurance in Design/Development, Production, Installation, and 
Servicing.'' The CGMP/QS information collections will assist FDA 
inspections of manufacturers for compliance with QS requirements 
encompassing design, production, installation, and servicing processes.
    Section 820.20(a) through (e) requires management with executive 
responsibility to establish, maintain, and/or review the following 
topics: (1) The quality policy, (2) the organizational structure, (3) 
the quality plan, and (4) the quality system procedures of the 
organization. Section 820.22 requires the conduct and documentation of 
QS audits and re-audits. Section 820.25(b) requires the establishment 
of procedures to identify training needs and documentation of such 
training.
    Section 820.30(a)(1) and (b) through (j) requires, in respective 
order, the establishment, maintenance, and/or documentation of the 
following topics: (1) Procedures to control design of class III and 
class II devices and certain class I devices as listed therein; (2) 
plans for design and development activities and updates; (3) procedures 
identifying, documenting, and approving design input requirements; (4) 
procedures defining design output, including acceptance criteria, and 
documentation of approved records; (5) procedures for formal review of 
design results and documentation of results in the design history file 
(DHF); (6) procedures for verifying device design and documentation of 
results and approvals in the DHF; (7) procedures for validating device 
design, including documentation of results in the DHF; (8) procedures 
for translating device design into production specifications; (9) 
procedures for documenting, verifying, and validating approved design 
changes before implementation of changes; and (10) the records and 
references constituting the DHF for each type of device.
    Section 820.40 requires manufacturers to establish and maintain 
procedures controlling approval and distribution of required documents 
and document changes. Section 820.40(a) and (b) requires the 
establishment and maintenance of procedures for the review, approval, 
issuance, and documentation of required records (documents) and changes 
to those records.
    Section 820.50(a) and (b) requires the establishment and 
maintenance of procedures and requirements to ensure service and 
product quality, records of acceptable suppliers, and purchasing data 
describing specified requirements for products and services.
    Sections 820.60 and 820.65 require, respectively, the establishment 
and maintenance of procedures for identifying all products from receipt 
to distribution and for using control numbers to track surgical 
implants and life-sustaining or supporting devices and their 
components.
    Section 820.70(a) through (e), (g)(1) through (g)(3), (h), and (i) 
requires the establishment, maintenance, and/or documentation of the 
following topics: (1) Process control procedures; (2) procedures for 
verifying or validating changes to specification, method, process, or 
procedure; (3) procedures to control environmental conditions and 
inspection result records; (4) requirements for personnel hygiene; (5) 
procedures for preventing contamination of equipment and products; (6) 
equipment adjustment, cleaning, and maintenance schedules; (7) 
equipment inspection records; (8) equipment tolerance postings, 
procedures for utilizing manufacturing materials expected to have an 
adverse effect on product quality; and (9) validation protocols and 
validation records for computer software and software changes.
    Sections 820.72(a), (b)(1), and (b)(2); and 820.75(a) through (c) 
require, respectively, the establishment, maintenance, and/or 
documentation of the following topics: (1) Equipment calibration and 
inspection procedures; (2) national, international, or in-house 
calibration standards; (3) records that identify calibrated equipment 
and next calibration dates; (4) validation procedures and validation 
results for processes not verifiable by inspections and tests; (5) 
procedures for keeping validated processes within specified limits; (6) 
records for monitoring and controlling validated processes; and (7) 
records of the results of revalidation where necessitated by process 
changes or deviations.
    Sections 820.80(a) through (e) and 820.86, respectively, require 
the establishment, maintenance, and/or documentation of the following 
topics: (1) Procedures for incoming acceptance by inspection, test, or 
other verification; (2) procedures for ensuring that in process 
products meet specified requirements and the control of product until 
inspection and tests are completed; (3) procedures for, and records 
that show, incoming acceptance or rejection is conducted by 
inspections, tests or other verifications; (4) procedures for, and 
records that show, finished devices meet acceptance criteria and are 
not distributed until device master record (DMR) activities are 
completed; (5) records in the device history record (DHR) showing 
acceptance dates, results, and equipment used; and (6) the acceptance/
rejection identification of products from receipt to installation and 
servicing.
    Sections 820.90(a), (b)(1), and (b)(2) and 820.100 require, 
respectively, the establishment, maintenance and/or documentation of 
the following topics: (1) Procedures for identifying, recording, 
evaluating, and disposing of nonconforming product; (2) procedures for 
reviewing and recording concessions made for, and disposition of, 
nonconforming product; (3) procedures for reworking products, 
evaluating possible adverse rework effect and recording results in the 
DHR; (4) procedures and requirements for corrective and preventive 
actions, including analysis, investigation, identification and review 
of data, records, causes, and results; and (5) records for all 
corrective and preventive action activities.
    Section 820.100(a)(1) through (a)(7) states that procedures and 
requirements shall be established and maintained for corrective/
preventive actions, including the following: (1) Analysis of data from 
process, work, quality, servicing records, investigation of 
nonconformance causes; (2) identification of corrections and their 
effectiveness; (3) recording of changes made; and (4) appropriate 
distribution and managerial review of corrective and preventive action 
information. Section 820.120 states that manufacturers shall establish/
maintain procedures to control labeling storage/application; and 
examination/release for storage and use, and document those procedures.
    Sections 820.120(b) and (d); 820.130; 820.140; 820.150(a) and (b); 
820.160(a) and (b); and 820.170(a) and (b), respectively, require the 
establishment, maintenance, and/or documentation of the following 
topics: (1) Procedures for controlling and recording the storage, 
examination, release, and use of labeling; (2) the filing of labels/
labeling used in the DHR; (3) procedures for controlling product 
storage areas and receipt/dispatch authorizations; (4) procedures 
controlling the release of products for distribution; (5) distribution 
records that identify consignee, product, date, and control numbers; 
and (6) instructions, inspection and test procedures that are

[[Page 62147]]

made available, and the recording of results for devices requiring 
installation.
    Sections 820.180(b) and (c); 820.181(a) through (e); 820.184(a) 
through (f); and 820.186 require, respectively, the maintenance of 
records that are: (1) Retained at prescribed site(s), made readily 
available and accessible to FDA, and retained for the device's life 
expectancy or for 2 years; (2) contained or referenced in a DMR 
consisting of device, process, quality assurance, packaging and 
labeling, and installation, maintenance, and servicing specifications 
and procedures; (3) contained in a DHR and demonstrate the manufacture 
of each unit, lot, or batch of product in conformance with DMR and 
regulatory requirements include manufacturing and distribution dates, 
quantities, acceptance documents, labels and labeling, and control 
numbers; and (4) contained in a quality system record, consisting of 
references, documents, procedures, and activities not specific to 
particular devices.
    Sections 820.198(a) through (c); and 820.200(a) through (d), 
respectively, require the establishment, maintenance, and/or 
documentation of the following topics: (1) Complaint files and 
procedures for receiving, reviewing, and evaluating complaints; (2) 
complaint investigation records identifying the device, complainant, 
and relationship of the device to the incident; (3) complaint records 
that are reasonably accessible to the manufacturing site or at 
prescribed sites; (4) procedures for performing and verifying that 
device servicing requirements are met and that service reports 
involving complaints are processed as complaints; and (5) service 
reports that record the device, service activity, and test and 
inspection data.
    Section 820.250 requires the establishment and maintenance of 
procedures to identify valid statistical techniques necessary to verify 
process and product acceptability; and sampling plans, when used, which 
are written and based on valid statistical rationale; and procedures 
for ensuring adequate sampling methods.
    The CGMP/QS regulation added design and purchasing controls, 
modified previous critical device requirements, revised previous 
validation and other requirements, and harmonized device CGMP 
requirements with QS specifications in the international standard ``ISO 
9001: Quality Systems Model for Quality Assurance in Design/
Development, Production, Installation, and Servicing.'' The rule does 
not apply to manufacturers of components or parts of finished devices, 
or to manufacturers of human blood and blood components subject to 21 
CFR part 606. With respect to devices classified in class I, design 
control requirements apply only to class I devices listed in Sec.  
820.30(a)(2) of the regulation. The rule imposes burden upon: (1) 
Finished device manufacturer firms, which are subject to all 
recordkeeping requirements; (2) finished device contract manufacturers, 
specification developers; and (3) re-packer, re-labelers, and contract 
sterilizer firms, which are subject only to requirements applicable to 
their activities. In addition, remanufacturers of hospital single-use 
devices are now considered to have the same requirements as 
manufacturers in regard to the regulation.
    The establishment, maintenance, and/or documentation of procedures, 
records, and data required by the regulation assists FDA in determining 
whether firms are in compliance with CGMP requirements, which are 
intended to ensure that devices meet their design, production, 
labeling, installation, and servicing specifications and, thus are 
safe, effective, and suitable for their intended purpose. In 
particular, compliance with CGMP design control requirements should 
decrease the number of design-related device failures that have 
resulted in deaths and serious injuries.
    The CGMP/QS regulation applies to approximately 24,738 respondents. 
A query of the Agency's registration and listing database shows that 
approximately 13,294 domestic and 11,444 foreign establishments are 
respondents to this information collection.\1\ Respondents to this 
collection have no reporting activities, but must make required records 
available for review or copying during FDA inspection. Except for 
manufacturers, not every type of firm is subject to every CGMP/QS 
requirement. For example, all are subject to Quality Policy (Sec.  
820.20(a)), Document Control (Sec.  820.40), and other requirements, 
whereas only manufacturers and specification developers are subject to 
subpart C, Design Controls. The PRA burden placed on the 24,738 
establishments is an average burden.
---------------------------------------------------------------------------

    \1\ Based on fiscal year 2015 data.

                               Table 1--Estimated Annual Recordkeeping Burden \1\
----------------------------------------------------------------------------------------------------------------
                                                     Number of                    Average burden
     Activity/21 CFR section         Number of      records per    Total annual         per         Total hours
                                   recordkeepers   recordkeeper       records      recordkeeping
----------------------------------------------------------------------------------------------------------------
Quality policy--820.20(a).......          24,738               1          24,738               7         173,166
Organization--820.20(b).........          24,738               1          24,738               4          98,952
Management review--820.20(c)....          24,738               1          24,738               6         148,428
Quality planning--820.20(d).....          24,738               1          24,738              10         247,380
Quality system procedures--               24,738               1          24,738              10         247,380
 820.20(e)......................
Quality audit--820.22...........          24,738               1          24,738              33         816,354
Training--820.25(b).............          24,738               1          24,738              13         321,594
Design procedures--820.30(a)(1).          24,738               1          24,738               2          49,476
Design and development planning--         24,738               1          24,738               6         148,428
 820.30(b)......................
Design input--820.30(c).........          24,738               1          24,738               2          49,476
Design output--820.30(d)........          24,738               1          24,738               2          49,476
Design review--820.30(e)........          24,738               1          24,738              23         568,974
Design verification--820.30(f)..          24,738               1          24,738              37         915,306
Design validation--820.30(g)....          24,738               1          24,738              37         915,306
Design transfer--820.30(h)......          24,738               1          24,738               3          74,214
Design changes--820.30(i).......          24,738               1          24,738              17         420,546
Design history file--820.30(j)..          24,738               1          24,738               3          74,214
Document controls--820.40.......          24,738               1          24,738               9         222,642
Documentation approval and                24,738               1          24,738               2          49,476
 distribution and document
 changes--820.40(a) and (b).....

[[Page 62148]]

 
Purchasing controls--820.50(a)..          24,738               1          24,738              22         544,236
Purchasing data--820.50(b)......          24,738               1          24,738               6         148,428
Identification--820.60..........          24,738               1          24,738               1          24,738
Traceability--820.65............          24,738               1          24,738               1          24,738
Production and process controls--         24,738               1          24,738               2          49,476
 820.70(a)......................
Production and process changes            24,738               1          24,738               2          49,476
 and environmental control--
 820.70(b) and (c)..............
Personnel--820.70(d)............          24,738               1          24,738               3          74,214
Contamination control--820.70(e)          24,738               1          24,738               2          49,476
Equipment maintenance schedule,           24,738               1          24,738               1          24,738
 inspection, and adjustment--
 820.70(g)(1)-(g)(3)............
Manufacturing material--                  24,738               1          24,738               2          49,476
 820.70(h)......................
Automated processes--820.70(i)..          24,738               1          24,738               8         197,904
Control of inspection,                    24,738               1          24,738               5         123,690
 measuring, and test equipment--
 820.72(a)......................
Calibration procedures,                   24,738               1          24,738               1          24,738
 standards, and records--
 820.72(b)(1)-(b)(2)............
Process validation--820.75(a)...          24,738               1          24,738               3          74,214
Validated process parameters,             24,738               1          24,738               1          24,738
 monitoring, control methods,
 and data--820.75(b)............
Revalidation--820.75(c).........          24,738               1          24,738               1          24,738
Acceptance activities--820.80(a)-         24,738               1          24,738               5         123,690
 (e)............................
Acceptance status--820.86.......          24,738               1          24,738               1          24,738
Control of nonconforming                  24,738               1          24,738               5         123,690
 product--820.90(a).............
Nonconforming product review/             24,738               1          24,738               5         123,690
 disposition procedures and
 rework procedures--820.90(b)(1)-
 (b)(2).........................
Procedures for corrective/                24,738               1          24,738              12         296,856
 preventive actions--
 820.100(a)(1)-(a)(7)...........
Corrective/preventive                     24,738               1          24,738               1          24,738
 activities--820.100(b).........
Labeling procedures--820.120(b).          24,738               1          24,738               1          24,738
Labeling documentation--                  24,738               1          24,738               1          24,738
 820.120(d).....................
Device packaging--820.130.......          24,738               1          24,738               1          24,738
Handling--820.140...............          24,738               1          24,738               6         148,428
Storage--820.150(a) and (b).....          24,738               1          24,738               6         148,428
Distribution procedures and               24,738               1          24,738               1          24,738
 records--820.160(a) and (b)....
Installation--820.170...........          24,738               1          24,738               2          49,476
Record retention period--                 24,738               1          24,738               2          49,476
 820.180(b) and (c).............
Device master record--820.181...          24,738               1          24,738               1          24,738
Device history record--820.184..          24,738               1          24,738               1          24,738
Quality system record--820.186..          24,738               1          24,738               1          24,738
Complaint files--820.198(a),              24,738               1          24,738               5         123,690
 (c), and (g)...................
Servicing procedures and                  24,738               1          24,738               3          74,214
 reports--820.200(a) and (d)....
Statistical techniques                    24,738               1          24,738               1          24,738
 procedures and sampling plans--
 820.250........................
                                 -------------------------------------------------------------------------------
    Total.......................  ..............  ..............  ..............  ..............       8,410,920
----------------------------------------------------------------------------------------------------------------
\1\ There are no capital costs or operating and maintenance costs associated with this collection of
  information.


    Dated: September 1, 2016.
Leslie Kux,
Associate Commissioner for Policy.
[FR Doc. 2016-21553 Filed 9-7-16; 8:45 am]
 BILLING CODE 4164-01-P



                                                  62144                     Federal Register / Vol. 81, No. 174 / Thursday, September 8, 2016 / Notices

                                                  ACTION:   Notice.                                       FDA may not approve an ANDA that                      Book. The ‘‘Discontinued Drug Product
                                                                                                          does not refer to a listed drug.                      List’’ delineates, among other items,
                                                  SUMMARY:    The Food and Drug                              PREVACID IV (lansoprazole)                         drug products that have been
                                                  Administration (FDA or Agency) has                      intravenous injection, 30 mg/vial, is the             discontinued from marketing for reasons
                                                  determined that PREVACID IV                             subject of NDA 021566, held by Takeda                 other than safety or effectiveness.
                                                  (lansoprazole) intravenous injection, 30                Pharmaceuticals North America, Inc.,                  ANDAs that refer to PREVACID IV
                                                  milligrams (mg)/vial, was not                           and initially approved on May 27, 2004.               (lansoprazole) intravenous injection, 30
                                                  withdrawn from sale for reasons of                      The Indications and Usage section of the              mg/vial, may be approved by the
                                                  safety or effectiveness. This                           PREVACID IV labeling states the                       Agency as long as they meet all other
                                                  determination will allow FDA to                         following: ‘‘When patients are unable to              legal and regulatory requirements for
                                                  approve abbreviated new drug                            take the oral formulations, PREVACID                  the approval of ANDAs. If FDA
                                                  applications (ANDAs) for lansoprazole                   I.V. for Injection is indicated as an                 determines that labeling for this drug
                                                  intravenous injection, 30 mg/vial, if all               alternative for the short-term treatment              product should be revised to meet
                                                  other legal and regulatory requirements                 (up to 7 days) of all grades of erosive               current standards, the Agency will
                                                  are met.                                                esophagitis. Once the patient is able to              advise ANDA applicants to submit such
                                                  FOR FURTHER INFORMATION CONTACT:                        take medications orally, therapy can be               labeling.
                                                  Bronwen Blass, Center for Drug                          switched to an oral formulation of
                                                                                                                                                                  Dated: September 1, 2016.
                                                  Evaluation and Research, Food and                       PREVACID for a total of 6 to 8 weeks.
                                                                                                          The safety and efficacy of PREVACID                   Leslie Kux,
                                                  Drug Administration, 10903 New
                                                                                                          I.V. for Injection as an initial treatment            Associate Commissioner for Policy.
                                                  Hampshire Ave., Bldg. 51, Rm. 6224,
                                                  Silver Spring, MD 20993–0002, 301–                      of erosive esophagitis have not been                  [FR Doc. 2016–21551 Filed 9–7–16; 8:45 am]
                                                  796–5092, Bronwen.blass@fda.hhs.gov.                    demonstrated. Refer to full prescribing               BILLING CODE 4164–01–P
                                                                                                          information for the oral formulations of
                                                  SUPPLEMENTARY INFORMATION: In 1984,
                                                                                                          PREVACID.’’
                                                  Congress enacted the Drug Price                            In a letter dated February 5, 2007,                DEPARTMENT OF HEALTH AND
                                                  Competition and Patent Term                             Takeda Pharmaceuticals North America,                 HUMAN SERVICES
                                                  Restoration Act of 1984 (Pub. L. 98–417)                Inc. notified FDA that PREVACID IV
                                                  (the 1984 amendments), which                            (lansoprazole) intravenous injection, 30              Food and Drug Administration
                                                  authorized the approval of duplicate                    mg/vial, was being discontinued, and
                                                  versions of drug products under an                      FDA moved the drug product to the                     [Docket No. FDA–2016–N–2544]
                                                  ANDA procedure. ANDA applicants                         ‘‘Discontinued Drug Product List’’
                                                  must, with certain exceptions, show that                section of the Orange Book.                           Agency Information Collection
                                                  the drug for which they are seeking                        Rose Zhao submitted a citizen                      Activities; Proposed Collection;
                                                  approval contains the same active                       petition dated March 18, 2016 (Docket                 Comment Request; Medical Device:
                                                  ingredient in the same strength and                     No. FDA–2016–P–1037), under 21 CFR                    Current Good Manufacturing Practice
                                                  dosage form as the ‘‘listed drug,’’ which               10.30, requesting that the Agency                     Quality System Regulations
                                                  is a version of the drug that was                       determine whether PREVACID IV
                                                  previously approved. ANDA applicants                                                                          AGENCY:    Food and Drug Administration,
                                                                                                          (lansoprazole) intravenous injection, 30              HHS.
                                                  do not have to repeat the extensive                     mg/vial, was withdrawn from sale for
                                                  clinical testing otherwise necessary to                 reasons of safety or effectiveness.                   ACTION:   Notice.
                                                  gain approval of a new drug application                    After considering the citizen petition             SUMMARY:  The Food and Drug
                                                  (NDA).                                                  and reviewing Agency records and                      Administration (FDA) is announcing an
                                                     The 1984 amendments include what                     based on the information we have at this              opportunity for public comment on the
                                                  is now section 505(j)(7) of the Federal                 time, FDA has determined under                        proposed collection of certain
                                                  Food, Drug, and Cosmetic Act (21 U.S.C.                 § 314.161 that PREVACID IV                            information by the Agency. Under the
                                                  355(j)(7)), which requires FDA to                       (lansoprazole) intravenous injection, 30              Paperwork Reduction Act of 1995 (the
                                                  publish a list of all approved drugs.                   mg/vial, was not withdrawn for reasons                PRA), Federal Agencies are required to
                                                  FDA publishes this list as part of the                  of safety or effectiveness. The petitioner            publish notice in the Federal Register
                                                  ‘‘Approved Drug Products With                           has identified no data or other                       concerning each proposed collection of
                                                  Therapeutic Equivalence Evaluations,’’                  information suggesting that PREVACID                  information including each proposed
                                                  which is known generally as the                         IV (lansoprazole) intravenous injection,              extension of an existing collection of
                                                  ‘‘Orange Book.’’ Under FDA regulations,                 30 mg/vial, was withdrawn for reasons                 information, and to allow 60 days for
                                                  drugs are removed from the list if the                  of safety or effectiveness. We have                   public comment in response to the
                                                  Agency withdraws or suspends                            carefully reviewed our files for records              notice. This notice solicits comments on
                                                  approval of the drug’s NDA or ANDA                      concerning the withdrawal of                          recordkeeping requirements related to
                                                  for reasons of safety or effectiveness or               PREVACID IV (lansoprazole)                            the medical devices current good
                                                  if FDA determines that the listed drug                  intravenous injection, 30 mg/vial, from               manufacturing practice (CGMP) quality
                                                  was withdrawn from sale for reasons of                  sale. We have also independently                      system (QS) regulation (CGMP/QS
                                                  safety or effectiveness (21 CFR 314.162).               evaluated relevant literature and data                regulation).
                                                     A person may petition the Agency to                  for possible postmarketing adverse
                                                  determine, or the Agency may                            events. We have reviewed the available                DATES:  Submit either electronic or
                                                  determine on its own initiative, whether                evidence and determined that this drug                written comments on the collection of
mstockstill on DSK3G9T082PROD with NOTICES




                                                  a listed drug was withdrawn from sale                   product was not withdrawn from sale                   information by November 7, 2016.
                                                  for reasons of safety or effectiveness.                 for reasons of safety or effectiveness.               ADDRESSES: You may submit comments
                                                  This determination may be made at any                      Accordingly, the Agency will                       as follows:
                                                  time after the drug has been withdrawn                  continue to list PREVACID IV
                                                  from sale, but must be made prior to                    (lansoprazole) intravenous injection, 30              Electronic Submissions
                                                  approving an ANDA that refers to the                    mg/vial, in the ‘‘Discontinued Drug                     Submit electronic comments in the
                                                  listed drug (§ 314.161 (21 CFR 314.161)).               Product List’’ section of the Orange                  following way:


                                             VerDate Sep<11>2014   19:34 Sep 07, 2016   Jkt 238001   PO 00000   Frm 00073   Fmt 4703   Sfmt 4703   E:\FR\FM\08SEN1.SGM   08SEN1


                                                                            Federal Register / Vol. 81, No. 174 / Thursday, September 8, 2016 / Notices                                            62145

                                                    • Federal eRulemaking Portal: http://                 ‘‘THIS DOCUMENT CONTAINS                              of the proposed collection of
                                                  www.regulations.gov. Follow the                         CONFIDENTIAL INFORMATION.’’ The                       information set forth in this document.
                                                  instructions for submitting comments.                   Agency will review this copy, including                  With respect to the following
                                                  Comments submitted electronically,                      the claimed confidential information, in              collection of information, FDA invites
                                                  including attachments, to http://                       its consideration of comments. The                    comments on these topics: (1) Whether
                                                  www.regulations.gov will be posted to                   second copy, which will have the                      the proposed collection of information
                                                  the docket unchanged. Because your                      claimed confidential information                      is necessary for the proper performance
                                                  comment will be made public, you are                    redacted/blacked out, will be available               of FDA’s functions, including whether
                                                  solely responsible for ensuring that your               for public viewing and posted on http://              the information will have practical
                                                  comment does not include any                            www.regulations.gov. Submit both                      utility; (2) the accuracy of FDA’s
                                                  confidential information that you or a                  copies to the Division of Dockets                     estimate of the burden of the proposed
                                                  third party may not wish to be posted,                  Management. If you do not wish your                   collection of information, including the
                                                  such as medical information, your or                    name and contact information to be                    validity of the methodology and
                                                  anyone else’s Social Security number, or                made publicly available, you can                      assumptions used; (3) ways to enhance
                                                  confidential business information, such                 provide this information on the cover                 the quality, utility, and clarity of the
                                                  as a manufacturing process. Please note                 sheet and not in the body of your                     information to be collected; and (4)
                                                  that if you include your name, contact                  comments and you must identify this                   ways to minimize the burden of the
                                                  information, or other information that                  information as ‘‘confidential.’’ Any                  collection of information on
                                                  identifies you in the body of your                      information marked as ‘‘confidential’’                respondents, including through the use
                                                  comments, that information will be                      will not be disclosed except in                       of automated collection techniques,
                                                  posted on http://www.regulations.gov.                   accordance with 21 CFR 10.20 and other                when appropriate, and other forms of
                                                    • If you want to submit a comment                     applicable disclosure law. For more                   information technology.
                                                  with confidential information that you                  information about FDA’s posting of                    Medical Devices: Current Good
                                                  do not wish to be made available to the                 comments to public dockets, see 80 FR                 Manufacturing Practice Quality System
                                                  public, submit the comment as a                         56469, September 18, 2015, or access                  Regulation—21 CFR Part 820—OMB
                                                  written/paper submission and in the                     the information at: http://www.fda.gov/               Control Number 0910–0073—Extension
                                                  manner detailed (see ‘‘Written/Paper                    regulatoryinformation/dockets/
                                                  Submissions’’ and ‘‘Instructions’’).                                                                             Under section 520(f) of the Federal
                                                                                                          default.htm.
                                                                                                                                                                Food, Drug, and Cosmetic Act (the
                                                  Written/Paper Submissions                                  Docket: For access to the docket to                FD&C Act) (21 U.S.C. 360j(f)), the
                                                     Submit written/paper submissions as                  read background documents or the                      Secretary of the Department of Health
                                                  follows:                                                electronic and written/paper comments                 and Human Services has the authority
                                                     • Mail/Hand delivery/Courier (for                    received, go to http://                               to prescribe regulations requiring that
                                                  written/paper submissions): Division of                 www.regulations.gov and insert the                    the methods used in, and the facilities
                                                  Dockets Management (HFA–305), Food                      docket number, found in brackets in the               and controls used for, the manufacture,
                                                  and Drug Administration, 5630 Fishers                   heading of this document, into the                    preproduction design validation
                                                  Lane, Rm. 1061, Rockville, MD 20852.                    ‘‘Search’’ box and follow the prompts                 (including a process to assess the
                                                     • For written/paper comments                         and/or go to the Division of Dockets                  performance of a device, but not
                                                  submitted to the Division of Dockets                    Management, 5630 Fishers Lane, Rm.                    including an evaluation of the safety
                                                  Management, FDA will post your                          1061, Rockville, MD 20852.                            and effectiveness of a device), packing,
                                                  comment, as well as any attachments,                    FOR FURTHER INFORMATION CONTACT:    FDA               storage, and installation of a device
                                                  except for information submitted,                       PRA Staff, Office of Operations, Food                 conform to CGMP, as described in such
                                                  marked and identified, as confidential,                 and Drug Administration, 8455                         regulations, to assure that the device
                                                  if submitted as detailed in                             Colesville Rd., COLE–14526, Silver                    will be safe and effective and otherwise
                                                  ‘‘Instructions.’’                                       Spring, MD 20993–0002, PRAStaff@                      in compliance with the FD&C Act.
                                                     Instructions: All submissions received               fda.hhs.gov.                                             The CGMP/QS regulation
                                                  must include the Docket No. FDA–                                                                              implementing authority provided by
                                                  2016–N–2544 for ‘‘Agency Information                    SUPPLEMENTARY INFORMATION:      Under the             this statutory provision is found under
                                                  Collection Activities; Proposed                         PRA (44 U.S.C. 3501–3520), Federal                    part 820 (21 CFR part 820) and sets forth
                                                  Collection; Comment Request; Medical                    Agencies must obtain approval from the                basic CGMP requirements governing the
                                                  Device: Current Good Manufacturing                      Office of Management and Budget                       design, manufacture, packing, labeling,
                                                  Practice Quality System Regulations.’’                  (OMB) for each collection of                          storage, installation, and servicing of all
                                                     Received comments will be placed in                  information they conduct or sponsor.                  finished medical devices intended for
                                                  the docket and, except for those                        ‘‘Collection of information’’ is defined              human use. The authority for this
                                                  submitted as ‘‘Confidential                             in 44 U.S.C. 3502(3) and 5 CFR                        regulation is covered under sections
                                                  Submissions,’’ publicly viewable at                     1320.3(c) and includes Agency requests                501, 502, 510, 513, 514, 515, 518, 519,
                                                  http://www.regulations.gov or at the                    or requirements that members of the                   520, 522, 701, 704, 801, and 803 of the
                                                  Division of Dockets Management                          public submit reports, keep records, or               FD&C Act (21 U.S.C. 351, 352, 360,
                                                  between 9 a.m. and 4 p.m., Monday                       provide information to a third party.                 360c, 360d, 360e, 360h, 360i, 360j, 360l,
                                                  through Friday.                                         Section 3506(c)(2)(A) of the PRA (44                  371, 374, 381, and 383). The CGMP/QS
                                                     • Confidential Submissions—To                        U.S.C. 3506(c)(2)(A)) requires Federal                regulation includes requirements for
                                                  submit a comment with confidential                      Agencies to provide a 60-day notice in                purchasing and service controls,
mstockstill on DSK3G9T082PROD with NOTICES




                                                  information that you do not wish to be                  the Federal Register concerning each                  clarifies recordkeeping requirements for
                                                  made publicly available, submit your                    proposed collection of information,                   device failure and complaint
                                                  comments only as a written/paper                        including each proposed extension of an               investigations, clarifies requirements for
                                                  submission. You should submit two                       existing collection of information,                   verifying/validating production
                                                  copies total. One copy will include the                 before submitting the collection to OMB               processes and process or product
                                                  information you claim to be confidential                for approval. To comply with this                     changes, and clarifies requirements for
                                                  with a heading or cover note that states                requirement, FDA is publishing notice                 product acceptance activities quality


                                             VerDate Sep<11>2014   19:34 Sep 07, 2016   Jkt 238001   PO 00000   Frm 00074   Fmt 4703   Sfmt 4703   E:\FR\FM\08SEN1.SGM   08SEN1


                                                  62146                     Federal Register / Vol. 81, No. 174 / Thursday, September 8, 2016 / Notices

                                                  data evaluations and corrections of                        Section 820.50(a) and (b) requires the             or rejection is conducted by inspections,
                                                  nonconforming product/quality                           establishment and maintenance of                      tests or other verifications; (4)
                                                  problems.                                               procedures and requirements to ensure                 procedures for, and records that show,
                                                     Requirements are compatible with                     service and product quality, records of               finished devices meet acceptance
                                                  specifications in the international                     acceptable suppliers, and purchasing                  criteria and are not distributed until
                                                  standards ‘‘ISO 9001: Quality Systems                   data describing specified requirements                device master record (DMR) activities
                                                  Model for Quality Assurance in Design/                  for products and services.                            are completed; (5) records in the device
                                                  Development, Production, Installation,                     Sections 820.60 and 820.65 require,                history record (DHR) showing
                                                  and Servicing.’’ The CGMP/QS                            respectively, the establishment and                   acceptance dates, results, and
                                                  information collections will assist FDA                 maintenance of procedures for                         equipment used; and (6) the acceptance/
                                                  inspections of manufacturers for                        identifying all products from receipt to              rejection identification of products from
                                                  compliance with QS requirements                         distribution and for using control                    receipt to installation and servicing.
                                                  encompassing design, production,                        numbers to track surgical implants and                   Sections 820.90(a), (b)(1), and (b)(2)
                                                  installation, and servicing processes.                  life-sustaining or supporting devices                 and 820.100 require, respectively, the
                                                     Section 820.20(a) through (e) requires               and their components.                                 establishment, maintenance and/or
                                                  management with executive                                  Section 820.70(a) through (e), (g)(1)              documentation of the following topics:
                                                  responsibility to establish, maintain,                  through (g)(3), (h), and (i) requires the             (1) Procedures for identifying,
                                                  and/or review the following topics: (1)                 establishment, maintenance, and/or                    recording, evaluating, and disposing of
                                                  The quality policy, (2) the                             documentation of the following topics:                nonconforming product; (2) procedures
                                                  organizational structure, (3) the quality               (1) Process control procedures; (2)                   for reviewing and recording concessions
                                                  plan, and (4) the quality system                        procedures for verifying or validating                made for, and disposition of,
                                                  procedures of the organization. Section                 changes to specification, method,                     nonconforming product; (3) procedures
                                                  820.22 requires the conduct and                         process, or procedure; (3) procedures to              for reworking products, evaluating
                                                  documentation of QS audits and re-                      control environmental conditions and                  possible adverse rework effect and
                                                  audits. Section 820.25(b) requires the                  inspection result records; (4)                        recording results in the DHR; (4)
                                                  establishment of procedures to identify                 requirements for personnel hygiene; (5)
                                                                                                                                                                procedures and requirements for
                                                  training needs and documentation of                     procedures for preventing
                                                                                                                                                                corrective and preventive actions,
                                                  such training.                                          contamination of equipment and
                                                     Section 820.30(a)(1) and (b) through                                                                       including analysis, investigation,
                                                                                                          products; (6) equipment adjustment,
                                                  (j) requires, in respective order, the                                                                        identification and review of data,
                                                                                                          cleaning, and maintenance schedules;
                                                  establishment, maintenance, and/or                                                                            records, causes, and results; and (5)
                                                                                                          (7) equipment inspection records; (8)
                                                  documentation of the following topics:                                                                        records for all corrective and preventive
                                                                                                          equipment tolerance postings,
                                                  (1) Procedures to control design of class                                                                     action activities.
                                                                                                          procedures for utilizing manufacturing
                                                  III and class II devices and certain class              materials expected to have an adverse                    Section 820.100(a)(1) through (a)(7)
                                                  I devices as listed therein; (2) plans for              effect on product quality; and (9)                    states that procedures and requirements
                                                  design and development activities and                   validation protocols and validation                   shall be established and maintained for
                                                  updates; (3) procedures identifying,                    records for computer software and                     corrective/preventive actions, including
                                                  documenting, and approving design                       software changes.                                     the following: (1) Analysis of data from
                                                  input requirements; (4) procedures                         Sections 820.72(a), (b)(1), and (b)(2);            process, work, quality, servicing
                                                  defining design output, including                       and 820.75(a) through (c) require,                    records, investigation of
                                                  acceptance criteria, and documentation                  respectively, the establishment,                      nonconformance causes; (2)
                                                  of approved records; (5) procedures for                 maintenance, and/or documentation of                  identification of corrections and their
                                                  formal review of design results and                     the following topics: (1) Equipment                   effectiveness; (3) recording of changes
                                                  documentation of results in the design                  calibration and inspection procedures;                made; and (4) appropriate distribution
                                                  history file (DHF); (6) procedures for                  (2) national, international, or in-house              and managerial review of corrective and
                                                  verifying device design and                             calibration standards; (3) records that               preventive action information. Section
                                                  documentation of results and approvals                  identify calibrated equipment and next                820.120 states that manufacturers shall
                                                  in the DHF; (7) procedures for validating               calibration dates; (4) validation                     establish/maintain procedures to control
                                                  device design, including documentation                  procedures and validation results for                 labeling storage/application; and
                                                  of results in the DHF; (8) procedures for               processes not verifiable by inspections               examination/release for storage and use,
                                                  translating device design into                          and tests; (5) procedures for keeping                 and document those procedures.
                                                  production specifications; (9)                          validated processes within specified                     Sections 820.120(b) and (d); 820.130;
                                                  procedures for documenting, verifying,                  limits; (6) records for monitoring and                820.140; 820.150(a) and (b); 820.160(a)
                                                  and validating approved design changes                  controlling validated processes; and (7)              and (b); and 820.170(a) and (b),
                                                  before implementation of changes; and                   records of the results of revalidation                respectively, require the establishment,
                                                  (10) the records and references                         where necessitated by process changes                 maintenance, and/or documentation of
                                                  constituting the DHF for each type of                   or deviations.                                        the following topics: (1) Procedures for
                                                  device.                                                    Sections 820.80(a) through (e) and                 controlling and recording the storage,
                                                     Section 820.40 requires manufacturers                820.86, respectively, require the                     examination, release, and use of
                                                  to establish and maintain procedures                    establishment, maintenance, and/or                    labeling; (2) the filing of labels/labeling
                                                  controlling approval and distribution of                documentation of the following topics:                used in the DHR; (3) procedures for
                                                  required documents and document                         (1) Procedures for incoming acceptance                controlling product storage areas and
mstockstill on DSK3G9T082PROD with NOTICES




                                                  changes. Section 820.40(a) and (b)                      by inspection, test, or other verification;           receipt/dispatch authorizations; (4)
                                                  requires the establishment and                          (2) procedures for ensuring that in                   procedures controlling the release of
                                                  maintenance of procedures for the                       process products meet specified                       products for distribution; (5)
                                                  review, approval, issuance, and                         requirements and the control of product               distribution records that identify
                                                  documentation of required records                       until inspection and tests are                        consignee, product, date, and control
                                                  (documents) and changes to those                        completed; (3) procedures for, and                    numbers; and (6) instructions,
                                                  records.                                                records that show, incoming acceptance                inspection and test procedures that are


                                             VerDate Sep<11>2014   19:34 Sep 07, 2016   Jkt 238001   PO 00000   Frm 00075   Fmt 4703   Sfmt 4703   E:\FR\FM\08SEN1.SGM   08SEN1


                                                                                  Federal Register / Vol. 81, No. 174 / Thursday, September 8, 2016 / Notices                                                    62147

                                                  made available, and the recording of                               processed as complaints; and (5) service              considered to have the same
                                                  results for devices requiring installation.                        reports that record the device, service               requirements as manufacturers in regard
                                                     Sections 820.180(b) and (c);                                    activity, and test and inspection data.               to the regulation.
                                                  820.181(a) through (e); 820.184(a)                                    Section 820.250 requires the                          The establishment, maintenance, and/
                                                  through (f); and 820.186 require,                                  establishment and maintenance of
                                                                                                                                                                           or documentation of procedures,
                                                  respectively, the maintenance of records                           procedures to identify valid statistical
                                                                                                                                                                           records, and data required by the
                                                  that are: (1) Retained at prescribed                               techniques necessary to verify process
                                                                                                                                                                           regulation assists FDA in determining
                                                  site(s), made readily available and                                and product acceptability; and sampling
                                                                                                                                                                           whether firms are in compliance with
                                                  accessible to FDA, and retained for the                            plans, when used, which are written
                                                                                                                                                                           CGMP requirements, which are
                                                  device’s life expectancy or for 2 years;                           and based on valid statistical rationale;
                                                                                                                                                                           intended to ensure that devices meet
                                                  (2) contained or referenced in a DMR                               and procedures for ensuring adequate
                                                  consisting of device, process, quality                             sampling methods.                                     their design, production, labeling,
                                                  assurance, packaging and labeling, and                                The CGMP/QS regulation added                       installation, and servicing specifications
                                                  installation, maintenance, and servicing                           design and purchasing controls,                       and, thus are safe, effective, and suitable
                                                  specifications and procedures; (3)                                 modified previous critical device                     for their intended purpose. In particular,
                                                  contained in a DHR and demonstrate the                             requirements, revised previous                        compliance with CGMP design control
                                                  manufacture of each unit, lot, or batch                            validation and other requirements, and                requirements should decrease the
                                                  of product in conformance with DMR                                 harmonized device CGMP requirements                   number of design-related device failures
                                                  and regulatory requirements include                                with QS specifications in the                         that have resulted in deaths and serious
                                                  manufacturing and distribution dates,                              international standard ‘‘ISO 9001:                    injuries.
                                                  quantities, acceptance documents,                                  Quality Systems Model for Quality                        The CGMP/QS regulation applies to
                                                  labels and labeling, and control                                   Assurance in Design/Development,                      approximately 24,738 respondents. A
                                                  numbers; and (4) contained in a quality                            Production, Installation, and Servicing.’’            query of the Agency’s registration and
                                                  system record, consisting of references,                           The rule does not apply to                            listing database shows that
                                                  documents, procedures, and activities                              manufacturers of components or parts of               approximately 13,294 domestic and
                                                  not specific to particular devices.                                finished devices, or to manufacturers of              11,444 foreign establishments are
                                                     Sections 820.198(a) through (c); and                            human blood and blood components                      respondents to this information
                                                  820.200(a) through (d), respectively,                              subject to 21 CFR part 606. With respect              collection.1 Respondents to this
                                                  require the establishment, maintenance,                            to devices classified in class I, design              collection have no reporting activities,
                                                  and/or documentation of the following                              control requirements apply only to class              but must make required records
                                                  topics: (1) Complaint files and                                    I devices listed in § 820.30(a)(2) of the             available for review or copying during
                                                  procedures for receiving, reviewing, and                           regulation. The rule imposes burden                   FDA inspection. Except for
                                                  evaluating complaints; (2) complaint                               upon: (1) Finished device manufacturer                manufacturers, not every type of firm is
                                                  investigation records identifying the                              firms, which are subject to all                       subject to every CGMP/QS requirement.
                                                  device, complainant, and relationship of                           recordkeeping requirements; (2)                       For example, all are subject to Quality
                                                  the device to the incident; (3) complaint                          finished device contract manufacturers,               Policy (§ 820.20(a)), Document Control
                                                  records that are reasonably accessible to                          specification developers; and (3) re-                 (§ 820.40), and other requirements,
                                                  the manufacturing site or at prescribed                            packer, re-labelers, and contract                     whereas only manufacturers and
                                                  sites; (4) procedures for performing and                           sterilizer firms, which are subject only              specification developers are subject to
                                                  verifying that device servicing                                    to requirements applicable to their                   subpart C, Design Controls. The PRA
                                                  requirements are met and that service                              activities. In addition, remanufacturers              burden placed on the 24,738
                                                  reports involving complaints are                                   of hospital single-use devices are now                establishments is an average burden.

                                                                                                     TABLE 1—ESTIMATED ANNUAL RECORDKEEPING BURDEN 1
                                                                                                                                                         Number of                              Average
                                                                                                                                       Number of                          Total annual
                                                                          Activity/21 CFR section                                                       records per                            burden per    Total hours
                                                                                                                                     recordkeepers                          records
                                                                                                                                                       recordkeeper                          recordkeeping

                                                  Quality policy—820.20(a) .....................................................            24,738                   1          24,738                   7       173,166
                                                  Organization—820.20(b) ......................................................             24,738                   1          24,738                   4        98,952
                                                  Management review—820.20(c) ..........................................                    24,738                   1          24,738                   6       148,428
                                                  Quality planning—820.20(d) ................................................               24,738                   1          24,738                  10       247,380
                                                  Quality system procedures—820.20(e) ...............................                       24,738                   1          24,738                  10       247,380
                                                  Quality audit—820.22 ..........................................................           24,738                   1          24,738                  33       816,354
                                                  Training—820.25(b) .............................................................          24,738                   1          24,738                  13       321,594
                                                  Design procedures—820.30(a)(1) .......................................                    24,738                   1          24,738                   2        49,476
                                                  Design and development planning—820.30(b) ...................                             24,738                   1          24,738                   6       148,428
                                                  Design input—820.30(c) ......................................................             24,738                   1          24,738                   2        49,476
                                                  Design output—820.30(d) ....................................................              24,738                   1          24,738                   2        49,476
                                                  Design review—820.30(e) ...................................................               24,738                   1          24,738                  23       568,974
                                                  Design verification—820.30(f) ..............................................              24,738                   1          24,738                  37       915,306
                                                  Design validation—820.30(g) ...............................................               24,738                   1          24,738                  37       915,306
mstockstill on DSK3G9T082PROD with NOTICES




                                                  Design transfer—820.30(h) ..................................................              24,738                   1          24,738                   3        74,214
                                                  Design changes—820.30(i) .................................................                24,738                   1          24,738                  17       420,546
                                                  Design history file—820.30(j) ...............................................             24,738                   1          24,738                   3        74,214
                                                  Document controls—820.40 ................................................                 24,738                   1          24,738                   9       222,642
                                                  Documentation approval and distribution and document
                                                    changes—820.40(a) and (b) ............................................                  24,738                   1          24,738                   2         49,476

                                                    1 Based   on fiscal year 2015 data.



                                             VerDate Sep<11>2014      19:34 Sep 07, 2016       Jkt 238001    PO 00000      Frm 00076   Fmt 4703   Sfmt 4703   E:\FR\FM\08SEN1.SGM   08SEN1


                                                  62148                               Federal Register / Vol. 81, No. 174 / Thursday, September 8, 2016 / Notices

                                                                                               TABLE 1—ESTIMATED ANNUAL RECORDKEEPING BURDEN 1—Continued
                                                                                                                                                                             Number of                                             Average
                                                                                                                                                 Number of                                            Total annual
                                                                            Activity/21 CFR section                                                                         records per                                           burden per               Total hours
                                                                                                                                               recordkeepers                                            records
                                                                                                                                                                           recordkeeper                                         recordkeeping

                                                  Purchasing controls—820.50(a) ..........................................                                 24,738                              1                 24,738                             22         544,236
                                                  Purchasing data—820.50(b) ................................................                               24,738                              1                 24,738                              6         148,428
                                                  Identification—820.60 ..........................................................                         24,738                              1                 24,738                              1          24,738
                                                  Traceability—820.65 ............................................................                         24,738                              1                 24,738                              1          24,738
                                                  Production and process controls—820.70(a) ......................                                         24,738                              1                 24,738                              2          49,476
                                                  Production and process changes and environmental con-
                                                    trol—820.70(b) and (c) .....................................................                           24,738                              1                 24,738                              2           49,476
                                                  Personnel—820.70(d) ..........................................................                           24,738                              1                 24,738                              3           74,214
                                                  Contamination control—820.70(e) .......................................                                  24,738                              1                 24,738                              2           49,476
                                                  Equipment maintenance schedule, inspection, and adjust-
                                                    ment—820.70(g)(1)–(g)(3) ...............................................                               24,738                               1                 24,738                              1         24,738
                                                  Manufacturing material—820.70(h) .....................................                                   24,738                               1                 24,738                              2         49,476
                                                  Automated processes—820.70(i) ........................................                                   24,738                               1                 24,738                              8        197,904
                                                  Control of inspection, measuring, and test equipment—
                                                    820.72(a) ..........................................................................                    24,738                              1                 24,738                             5         123,690
                                                  Calibration procedures, standards, and records—
                                                    820.72(b)(1)–(b)(2) ...........................................................                         24,738                             1                  24,738                              1          24,738
                                                  Process validation—820.75(a) .............................................                                24,738                             1                  24,738                              3          74,214
                                                  Validated process parameters, monitoring, control meth-
                                                    ods, and data—820.75(b) ................................................                                24,738                             1                  24,738                              1         24,738
                                                  Revalidation—820.75(c) .......................................................                            24,738                             1                  24,738                              1         24,738
                                                  Acceptance activities—820.80(a)–(e) ..................................                                    24,738                             1                  24,738                              5        123,690
                                                  Acceptance status—820.86 .................................................                                24,738                             1                  24,738                              1         24,738
                                                  Control of nonconforming product—820.90(a) ....................                                           24,738                             1                  24,738                              5        123,690
                                                  Nonconforming product review/disposition procedures and
                                                    rework procedures—820.90(b)(1)–(b)(2) .........................                                         24,738                              1                24,738                               5        123,690
                                                  Procedures         for         corrective/preventive                    actions—
                                                    820.100(a)(1)–(a)(7) .........................................................                          24,738                             1                  24,738                            12         296,856
                                                  Corrective/preventive activities—820.100(b) .......................                                       24,738                             1                  24,738                             1          24,738
                                                  Labeling procedures—820.120(b) .......................................                                    24,738                             1                  24,738                             1          24,738
                                                  Labeling documentation—820.120(d) ..................................                                      24,738                             1                  24,738                             1          24,738
                                                  Device packaging—820.130 ................................................                                 24,738                             1                  24,738                             1          24,738
                                                  Handling—820.140 ..............................................................                           24,738                             1                  24,738                             6         148,428
                                                  Storage—820.150(a) and (b) ...............................................                                24,738                             1                  24,738                             6         148,428
                                                  Distribution procedures and records—820.160(a) and (b) ..                                                 24,738                             1                  24,738                             1          24,738
                                                  Installation—820.170 ...........................................................                          24,738                             1                  24,738                             2          49,476
                                                  Record retention period—820.180(b) and (c) ......................                                         24,738                             1                  24,738                             2          49,476
                                                  Device master record—820.181 ..........................................                                   24,738                             1                  24,738                             1          24,738
                                                  Device history record—820.184 ..........................................                                  24,738                             1                  24,738                             1          24,738
                                                  Quality system record—820.186 .........................................                                   24,738                             1                  24,738                             1          24,738
                                                  Complaint files—820.198(a), (c), and (g) ............................                                     24,738                             1                  24,738                             5         123,690
                                                  Servicing procedures and reports—820.200(a) and (d) ......                                                24,738                             1                  24,738                             3          74,214
                                                  Statistical techniques procedures and sampling plans—
                                                    820.250 .............................................................................                   24,738                              1                 24,738                             1           24,738

                                                        Total ..............................................................................   ........................   ........................   ........................   ........................      8,410,920
                                                     1 There    are no capital costs or operating and maintenance costs associated with this collection of information.


                                                    Dated: September 1, 2016.                                                ACTION:Notice; renewal of advisory                                          Commissioner formally determines that
                                                  Leslie Kux,                                                                committee.                                                                  renewal is in the public interest.
                                                  Associate Commissioner for Policy.                                                                                                                     FOR FURTHER INFORMATION CONTACT:
                                                  [FR Doc. 2016–21553 Filed 9–7–16; 8:45 am]                                 SUMMARY:   The Food and Drug                                                Lauren Tesh, Center for Drug Evaluation
                                                  BILLING CODE 4164–01–P
                                                                                                                             Administration (FDA) is announcing the                                      and Research, Food and Drug
                                                                                                                             renewal of the Oncologic Drugs                                              Administration, 10903 New Hampshire
                                                                                                                             Advisory Committee by the                                                   Ave., Bldg. 31, Rm. 2417, Silver Spring,
                                                  DEPARTMENT OF HEALTH AND                                                   Commissioner of Food and Drugs (the                                         MD 20993–0002, 301–796–9001,
                                                  HUMAN SERVICES                                                             Commissioner). The Commissioner has                                         ODAC@fda.hhs.gov.
                                                                                                                             determined that it is in the public                                         SUPPLEMENTARY INFORMATION: Under 41
                                                  Food and Drug Administration                                               interest to renew the Oncologic Drugs                                       CFR 102–3.65 and approval by the
mstockstill on DSK3G9T082PROD with NOTICES




                                                                                                                             Advisory Committee for an additional 2                                      Department of Health and Human
                                                  [Docket No. FDA–2016–N–0001]                                               years beyond the charter expiration                                         Services pursuant to 45 CFR part 11 and
                                                                                                                             date. The new charter will be in effect                                     by the General Services Administration,
                                                  Advisory Committee; Oncologic Drugs                                        until September 1, 2018.
                                                  Advisory Committee, Renewal                                                                                                                            FDA is announcing the renewal of the
                                                                                                                             DATES: Authority for the Oncologic                                          Oncologic Drugs Advisory Committee.
                                                  AGENCY:       Food and Drug Administration,                                Drugs Advisory Committee will expire                                        The committee is a discretionary
                                                  HHS.                                                                       on September 1, 2016, unless the                                            Federal advisory committee established


                                             VerDate Sep<11>2014        19:34 Sep 07, 2016          Jkt 238001       PO 00000       Frm 00077       Fmt 4703        Sfmt 4703      E:\FR\FM\08SEN1.SGM               08SEN1



Document Created: 2018-02-09 13:12:29
Document Modified: 2018-02-09 13:12:29
CategoryRegulatory Information
CollectionFederal Register
sudoc ClassAE 2.7:
GS 4.107:
AE 2.106:
PublisherOffice of the Federal Register, National Archives and Records Administration
SectionNotices
ActionNotice.
DatesSubmit either electronic or written comments on the collection of information by November 7, 2016.
ContactFDA PRA Staff, Office of Operations, Food and Drug Administration, 8455 Colesville Rd., COLE-14526, Silver Spring, MD 20993-0002, [email protected]
FR Citation81 FR 62144 